FDA Adverse Event Malfunction Summary report: N

AUTOSOFT 90

MDR report key: 21064877 · Received January 3, 2025

Report

Report Number
3003442380-2024-35810
Event Type
Malfunction
Date Received
January 3, 2025
Date of Event
November 15, 2024
Report Date
January 25, 2025
Manufacturer
UNOMEDICAL UM-D
Product Code
FPA
UDI-DI
05705244018129
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 01 - MDR 2083400 -MDR 3003442380-2024-35810. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: THE REFERENCE SAMPLES FOR THE LOT 6007373 HAVE ALREADY BEEN PREVIOUSLY TESTED IN THE PR 2062013 ON 22/JAN/2025. TEST RESULTS: THE REFERENCE SAMPLES WERE VISUALLY INSPECTED AND TESTED FOR AIR FLOW. ALL TEST RESULTS WERE WITHIN SPECIFICATIONS AS PER THE TEST REPORT TW PR. (B)(4) COMPLAINT TEST REPORT.PDF ATTACHED IN THIS RECORD. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6007373 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION VERSION 114 MANUFACTURED IN THE LINE 8, ON 03/JUN/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABSE ON 21/JAN/2025 AGAINST MALFUNCTION CODE "SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE." AND LOT 6007373 AND ANOTHER 22 COMPLAINTS HAVE BEEN REGISTERED IN TW FOR THE SAME LOT 6007373 AND MALFUNCTION CODE.

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR (B)(4)- DEVICE 4 OF 4.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN SPAIN. IT WAS REPORTED THAT THE PATIENT FACED FOUR INFUSION SETS CANNULA KINKED EVENTS ON (B)(6) 2024. EVENTS OCCURRED WITHIN 3 HOURS OF INSERTION. THE INSERTION SITE WAS AT ABDOMEN. PATIENT REPLACED INFUSION SETS AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. COMPANY DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
434168 AUTOSOFT 90 UNO INSET II 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL UM-D 1002817 6007373 05705244018129

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown