ACCESS® 2 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2011-01611
- Event Type
- Malfunction
- Date Received
- May 27, 2011
- Date of Event
- April 29, 2011
- Report Date
- April 29, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- MMI
- PMA / PMN Number
- K922823/A007
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
SAMPLE COLLECTION AND CENTRIFUGATION INFORMATION WERE NOT SUPPLIED. PER THE CUSTOMER, ACCUTNI QC WAS WITHIN THE CUSTOMER'S ESTABLISHED RANGES PRIOR TO THE EVENT. SPECIFIC QC DATA HAVE NOT BEEN SUPPLIED. SYSTEM CHECK INFORMATION HAS NOT BEEN SUPPLIED TO DATE. BCI FIELD SERVICE ENGINEER (FSE) ADDRESSED SEVERAL HARDWARE ISSUES BY CLEANING WASH WHEEL, PULLEY AND PINCH ROLLERS, AND BY REPLACING MIXER AND RV INCUBATOR BELT CLIPS, QUAD STEPPER MOTOR PCB BOARD, REBUILDING WASH PUMP WITH NEW BELT AND SEAL. FSE PERFORMED REQUIRED ALIGNMENTS AND PRIMED THE SYSTEM. FSE PERFORMED ROUTINE SYSTEM CHECK AND A HIGH SENSITIVITY SYSTEM CHECK. BOTH PASSED WITHIN INSTRUMENT SPECIFICATIONS. FSE PERFORMED QC FOR ACCUTNI WHICH PASSED. ALTHOUGH SEVERAL HARDWARE ISSUES WERE ADDRESSED, A DEFINITIVE ROOT CAUSE HAS NOT BEEN DETERMINED TO DATE FOR THIS EVENT.
A CUSTOMER REPORTED TO BECKMAN COULTER INC (BCI) THAT THE ACCESS 2 IMMUNOASSAY ANALYZER GENERATED A HIGHER THAN EXPECTED TROPONIN (ACCUTNI) RESULT FOR ONE (1) PATIENT. THE RESULT WAS NOT REPORTED OUT OF THE LAB. SUBSEQUENT TESTING PERFORMED ON AN ALTERNATE METHODOLOGY PRODUCED A LOWER RESULT THAT BETTER MATCHED OTHER CLINICAL CARDIAC TESTING. THE CUSTOMER DID NOT RECEIVE ANY REPORT OF PATIENT INJURY REQUIRING MEDICAL INTERVENTION OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS® 2 IMMUNOASSAY SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | MMI | BECKMAN COULTER INC. | ACCESS® 2 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |