FDA Adverse Event Malfunction Summary report: N

ACCESS® 2 IMMUNOASSAY SYSTEM

MDR report key: 2105850 · Received May 27, 2011

Report

Report Number
2122870-2011-01610
Event Type
Malfunction
Date Received
May 27, 2011
Date of Event
April 29, 2011
Report Date
April 29, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
MMI
PMA / PMN Number
K922823/A007
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SAMPLE COLLECTION AND CENTRIFUGATION INFORMATION WERE NOT SUPPLIED. PER THE CUSTOMER, ACCUTNI QC THAT SHOULD BE RESULTING NEAR 0.2 NG/ML WAS RESULTING AT > 35 NG/ML. SPECIFIC QC DATA HAS NOT BEEN SUPPLIED TO DATE. SYSTEM CHECK INFORMATION ALSO HAS NOT BEEN SUPPLIED TO DATE. BCI FIELD SERVICE ENGINEER (FSE) ADDRESSED SEVERAL HARDWARE ISSUES BY CLEANING WASH WHEEL, PULLEY AND PINCH ROLLERS, AND BY REPLACING MIXER AND RV INCUBATOR BELT CLIPS, QUAD STEPPER MOTOR PCB BOARD, REBUILDING WASH PUMP WITH NEW BELT AND SEAL. FSE PERFORMED REQUIRED ALIGNMENTS AND PRIMED THE SYSTEM. FSE PERFORMED ROUTINE SYSTEM CHECK AND A HIGH SENSITIVITY SYSTEM CHECK. BOTH PASSED WITHIN INSTRUMENT SPECIFICATIONS. FSE PERFORMED QC FOR ACCUTNI WHICH PASSED. ALTHOUGH SEVERAL HARDWARE ISSUES WERE ADDRESSED, A DEFINITIVE ROOT CAUSE HAS NOT BEEN DETERMINED TO DATE FOR THIS EVENT.

Description of Event or Problem · 1

A CUSTOMER REPORTED TO BECKMAN COULTER INC. (BCI) THAT THE ACCESS 2 IMMUNOASSAY ANALYZER GENERATED A HIGHER THAN EXPECTED TROPONIN (ACCUTNI) RESULT ABOVE THE AMI CUT-OFF FOR ONE (1) PATIENT . THE RESULT WAS NOT REPORTED OUT OF THE LAB. SUBSEQUENT TESTING PERFORMED ON AN ALTERNATE METHODOLOGY PRODUCED A LOWER RESULT THAT BETTER MATCHED OTHER CLINICAL CARDIAC TESTING. RESULTS ARE SHOWN. THE CUSTOMER DID NOT RECEIVE ANY REPORT OF PATIENT INJURY REQUIRING MEDICAL INTERVENTION OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED TO THIS EVENT.