FDA Adverse Event Malfunction Summary report: N

X-SERIES HUDSON HTQ RMER ATTACH

MDR report key: 21053290 · Received January 2, 2025

Report

Report Number
0008031000-2025-00001
Event Type
Malfunction
Date Received
January 2, 2025
Date of Event
December 5, 2024
Report Date
June 5, 2025
Manufacturer
ZIMMER SURGICAL SA
Product Code
GEY
UDI-DI
00889024521230
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4) G2 ¿ FOREIGN ¿ GERMANY. INVESTIGATION OF THIS INCIDENT IS CURRENTLY ONGOING. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4, B5, D9, G1, G3, G6, H1, H2, H3, H4, H6, H11. CORRECTED INFORMATION: D4 UDI NUMBER. REVIEW OF THE MOST RECENT REPAIR RECORD DETERMINED THE REPORTED EVENT COULD NOT BE DUPLICATED AS THE DEVICE PASSED THE DRILL LOCKING TEST. HOWEVER, THE LOCKING MECHANISM WAS DAMAGED. THE UNIT WAS SCRAPPED. REVIEW OF THE DEVICE HISTORY RECORD IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. COMPLAINTS ARE MONITORED THROUGH MONTHLY COMPLAINT REVIEW IN ORDER TO IDENTIFY POTENTIAL ADVERSE TRENDS. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. THE EVENT CANNOT BE CONFIRMED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

THE LOSS OF ATTACHMENT OF THE DRILL ADAPTER WAS REPORTED. NO CONSEQUENCES OR IMPACT ON THE PATIENT. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1536414 X-SERIES HUDSON HTQ RMER ATTACH SURGICAL POWER TOOL GEY ZIMMER SURGICAL SA 5014946 00889024521230

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown