FDA Adverse Event Malfunction Summary report: N

MOLLRING CUTTER TRANSECTION DEVICE

MDR report key: 2104796 · Received May 26, 2011

Report

Report Number
1220948-2011-00004
Event Type
Malfunction
Date Received
May 26, 2011
Report Date
April 15, 2011
Manufacturer
LEMAITRE VASCULAR
Product Code
DWX
PMA / PMN Number
K950813
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

WE HAVE RECEIVED THE DEVICE FOR EVALUATION AND WERE ABLE TO VERIFY THE FAILURE. THE TOP (PROXIMAL) RING WELDING EVALUATION DID NOT REVEAL ANY PROBLEMS DURING THE MANUFACTURING PROCESS - WELD WAS DONE CORRECTLY. THE ROOT CAUSE OF THE FAILURE IS INCONCLUSIVE. IT IS POSSIBLE THE EXCESSIVE FORCE AND/OR CALCIFIED VESSEL CONTRIBUTED TO THE FAILURE. PLEASE NOTE THAT WE DO NOT RECOMMEND THE USE OF MOLLRING CUTTER DEVICES FOR CUTTING CALCIFIED MATERIAL SINCE IT MAY DAMAGE THE DEVICE (PLEASE REFERENCE LEMAITRE VASCULAR, INC. IFU PRECAUTION SECTION FOR MORE DETAILS - SEE ATTACHMENT). PLEASE NOTE THAT THE PATIENT IS FINE AND THE CASE WAS DONE SUCCESSFULLY.

Description of Event or Problem · 1

THE MOLLRING DEVICE WAS USED FOR ENDARTERECTOMY PROCEDURE. THE DEVICE WAS PASSED TO THE DISTAL PORTION OF THE SFA, WHERE IT WAS INTENDED TO CUT THE PLAQUE. THE PLAQUE WAS CALCIFIED DISTALLY. WHEN DR (B)(6) PULLED THE WHITE HANDLE AWAY FROM THE BLUE TO ENGAGE THE CUTTER, THE TOP CUTTING RING DISENGAGED ITSELF PERPENDICULAR TO THE BOTTOM BLADE UNDER FLUORO. IT WAS ASSUMED THAT THERE WAS AN ISSUE WITH THE CUTTER. THE PHYSICIAN WAS UNABLE TO GET THE CUTTING RINGS TO GO BACK TOGETHER, BUT HE WAS ABLE TO REMOVE THE CORE. UPON INSPECTION, THE TOP RING WAS CAUGHT AROUND THE CORE. WHEN THE CORE WAS REMOVED FROM THE RING, IT WAS FOUND THAT THE TOP RING HAD ACTUALLY BROKEN AT THE WELD AWAY FROM ITS SHAFT AND HAD GOTTEN STUCK IN THE CORE. PATIENT IS FINE AND THE CASE WAS DONE SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MOLLRING CUTTER TRANSECTION DEVICE MOLLRING CUTTER DWX LEMAITRE VASCULAR 4200-42 MOL1055

Patients

Seq Age Sex Outcome Treatment
1