MOLLRING CUTTER TRANSECTION DEVICE
Report
- Report Number
- 1220948-2011-00004
- Event Type
- Malfunction
- Date Received
- May 26, 2011
- Report Date
- April 15, 2011
- Manufacturer
- LEMAITRE VASCULAR
- Product Code
- DWX
- PMA / PMN Number
- K950813
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PHYSICIAN
Narratives
WE HAVE RECEIVED THE DEVICE FOR EVALUATION AND WERE ABLE TO VERIFY THE FAILURE. THE TOP (PROXIMAL) RING WELDING EVALUATION DID NOT REVEAL ANY PROBLEMS DURING THE MANUFACTURING PROCESS - WELD WAS DONE CORRECTLY. THE ROOT CAUSE OF THE FAILURE IS INCONCLUSIVE. IT IS POSSIBLE THE EXCESSIVE FORCE AND/OR CALCIFIED VESSEL CONTRIBUTED TO THE FAILURE. PLEASE NOTE THAT WE DO NOT RECOMMEND THE USE OF MOLLRING CUTTER DEVICES FOR CUTTING CALCIFIED MATERIAL SINCE IT MAY DAMAGE THE DEVICE (PLEASE REFERENCE LEMAITRE VASCULAR, INC. IFU PRECAUTION SECTION FOR MORE DETAILS - SEE ATTACHMENT). PLEASE NOTE THAT THE PATIENT IS FINE AND THE CASE WAS DONE SUCCESSFULLY.
THE MOLLRING DEVICE WAS USED FOR ENDARTERECTOMY PROCEDURE. THE DEVICE WAS PASSED TO THE DISTAL PORTION OF THE SFA, WHERE IT WAS INTENDED TO CUT THE PLAQUE. THE PLAQUE WAS CALCIFIED DISTALLY. WHEN DR (B)(6) PULLED THE WHITE HANDLE AWAY FROM THE BLUE TO ENGAGE THE CUTTER, THE TOP CUTTING RING DISENGAGED ITSELF PERPENDICULAR TO THE BOTTOM BLADE UNDER FLUORO. IT WAS ASSUMED THAT THERE WAS AN ISSUE WITH THE CUTTER. THE PHYSICIAN WAS UNABLE TO GET THE CUTTING RINGS TO GO BACK TOGETHER, BUT HE WAS ABLE TO REMOVE THE CORE. UPON INSPECTION, THE TOP RING WAS CAUGHT AROUND THE CORE. WHEN THE CORE WAS REMOVED FROM THE RING, IT WAS FOUND THAT THE TOP RING HAD ACTUALLY BROKEN AT THE WELD AWAY FROM ITS SHAFT AND HAD GOTTEN STUCK IN THE CORE. PATIENT IS FINE AND THE CASE WAS DONE SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MOLLRING CUTTER TRANSECTION DEVICE | MOLLRING CUTTER | DWX | LEMAITRE VASCULAR | 4200-42 | MOL1055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |