FDA Adverse Event Injury Summary report: N

MINIMED QUICK-SET

MDR report key: 21046156 · Received December 31, 2024

Report

Report Number
3003442380-2024-35257
Event Type
Injury
Date Received
December 31, 2024
Date of Event
November 17, 2024
Report Date
May 20, 2025
Manufacturer
UNOMEDICAL UM-D
Product Code
FPA
UDI-DI
05705244017818
PMA / PMN Number
K160648
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED FOR THE MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION. THE BATCH 6001419 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. COMPLAINT INVESTIGATION: THE REFERENCE SAMPLES CANNOT BE TESTED BECAUSE THERE WAS NO SPECIFIC MALFUNCTION TO INVESTIGATE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6001419 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 75 PACKAGING IN THE MULTIVAC 14, ON 23/MAY/2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 19/MAY/2025 AGAINST MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION, HARM CODE SIGNS OF UNTREATED HYPOGLYCEMIA THAT PATIENT IS UNABLE TO SELF-MANAGE REQUIRING INTERVENTION BY AN HCP OR REQUIRES EMERGENCY ADVANCED LIFE SUPPORT TO PREVENT PERMANENT ORGAN DAMAGE AND LOT 6001419 AND NO OTHER COMPLAINTS HAVE BEEN REGISTERED IN DATABASE FOR THE SAME LOT NUMBER, HARM CODE AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, NO OTHER COMPLAINT RECEIVED ON THE LOT IN QUESTION, HARM CODE AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION OR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Additional Manufacturer Narrative · 0

E1 : PATIENT CITY : (B)(6). PATIENT COUNTRY : UNITED STATES.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED ON (B)(6) 2024 DUE TO HYPOGLYCEMIA. THE HOSPITALIZATION WAS LESS THAN 24 HOURS. THE BLOOD GLUCOSE LEVEL WAS 90MG/DL AT THE TIME OF EVENT AND THE PATIENT GOT TREATED WITH INTRAVENOUS DRIP AND WATER. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1253376 MINIMED QUICK-SET UNO QUICK-SET 80/9 SC1 MECA FPA UNOMEDICAL UM-D MMT-386A 6001419 05705244017818

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention| H