FDA Adverse Event Injury Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 2103780 · Received May 18, 2011

Report

Report Number
2032227-2011-01264
Event Type
Injury
Date Received
May 18, 2011
Date of Event
May 9, 2011
Report Date
May 10, 2011
Manufacturer
MEDTRONIC MINIMED
Product Code
LZG
PMA / PMN Number
K040676
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT HE WAS HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS, WITH A BLOOD GLUCOSE READING OF 400 MG/DL AND VOMITING. THE CUSTOMER HAD BEEN EXPERIENCING HIGH BLOOD GLUCOSE LEVELS FOR THE PAST DAY. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PUMP DID NOT PASS THE PRIME TEST. THE CUSTOMER REPORTED BENT CANNULAS AS WELL. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG MEDTRONIC MINIMED MMT-715LNAB

Patients

Seq Age Sex Outcome Treatment
1 65 YR Hospitalization SET. MMT-397, LOT 9202145| UNOMEDICAL QUICK-SET PARADIGM INSULIN INFUSION