QUATTRODE LEAD WIDE SPACED, 60 CM
Report
- Report Number
- 1627487-2011-00704
- Event Type
- Injury
- Date Received
- May 24, 2011
- Date of Event
- April 25, 2011
- Report Date
- April 25, 2011
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- K072462
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: THE DEVICE HISTORY AND STERILIZATION RECORDS REVIEWED WERE FOUND TO MEET SPECIFICATIONS AND NO ANOMALIES WERE FOUND. ONE OF THE THREE LEADS WAS RETURNED INCOMPLETE. AS SUCH, NO FUNCTIONAL TESTING COULD BE PERFORMED ON THIS DEVICE. WITH RESPECT TO THE OTHER LEADS, LEAD A PASSED ALL FUNCTIONAL TESTING; HOWEVER, LEAD B HAD BROKEN WIRES AT CHANNELS TWO AND FOUR AND THESE CHANNELS MEASURED OPEN IN CONTINUITY TESTING. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 4. REFERENCE MFR REPORT #S: 1627487-2011-00703, 1627487-2011-00705 AND 1627487-2011-00706. THE PT REC'D AN SCS SYSTEM FOR BACK PAIN INCLUDING AN IPG, FOUR PERCUTANEOUS LEADS (FROM TWO DIFFERENT LOTS) AND A LEAD EXTENSION. IT WAS REPORTED THAT THE PT'S SCS SYSTEM WAS EXPLANTED DUE TO ALLEGATIONS OF INEFFECTIVE STIMULATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | QUATTRODE LEAD WIDE SPACED, 60 CM | SPINAL CORD STIMULATION LEAD | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3166 | 174766 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Required Intervention |