FDA Adverse Event
Injury
Summary report: N
HAKIM VALVE
MDR report key: 2103709
·
Received May 23, 2011
Report
- Report Number
- 1226348-2011-00189
- Event Type
- Injury
- Date Received
- May 23, 2011
- Manufacturer
- CODMAN & SHURTLEFF, INC., MEDOS S.A.
- Product Code
- JXG
- PMA / PMN Number
- K974739
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CODMAN HAS REQUESTED RETURN OF THE VALVE FOR EVALUATION. HISTORICALLY, FOR COMPLAINTS OF THIS NATURE, EXAMINATIONS OF THE VALVES AND OR CATHETERS HAVE REVEALED THE ACCUMULATIONS OF BIOLOGICAL DEBRIS WITHIN THE DEVICE, WHICH HAVE RESULTED IN BLOCKAGES. REVIEW OF THE DEVICE HISTORY RECORDS HAVE CONFIRMED THAT THE VALVES HAVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR IF ANY FURTHER INFO REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CLOSED AT THIS TIME.
Description of Event or Problem · 1
AFFILIATE REPORTED THAT FLUID DID NOT FLOW THROUGH THE DEVICE. AS A RESULT THE DEVICE WAS REVISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HAKIM VALVE | SHUNT, CENTRAL NERVOUS SYSTEM & COMPS | JXG | CODMAN & SHURTLEFF, INC., MEDOS S.A. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Required Intervention |