CHPV
Report
- Report Number
- 1226348-2011-00191
- Event Type
- Injury
- Date Received
- May 23, 2011
- Date of Event
- April 22, 2011
- Manufacturer
- CODMAN & SHURTLEFF, INC., MEDOS S.A.
- Product Code
- JXG
- PMA / PMN Number
- UNK
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
HISTORICALLY, FOR COMPLAINTS OF THIS NATURE, EXAMINATIONS OF THE VALVES HAVE REVEALED THE PRESENCE OF BIOLOGICAL DEBRIS WITHIN THE DEVICES WHICH HAVE INTERFERED WITH THEIR PROGRAMMING AND/OR PRESSURE CONTROL MECHANISMS. THE BIOLOGICAL DEBRIS HAS CAUSED THE RETURNED VALVES TO FAIL THE PROGRAMMING AND/OR PRESSURE TESTS AND APPEARS TO HAVE CAUSED THE DIFFICULTIES EXPERIENCED BY THE CUSTOMERS. REVIEWS OF THE DEVICE HISTORY RECORDS HAVE CONFIRMED THAT THE VALVES HAVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR IF ANY FURTHER INFO REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CONSIDERED TO BE CLOSED AT THIS TIME.
AFFILIATE REPORTED AFTER THE MRI THE VALVE WAS NO LONGER PROGRAMMABLE. AS A RESULT THE DEVICE WAS REVISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CHPV | SHUNT, CENTRAL NERVOUS SYSTEM AND COMPS | JXG | CODMAN & SHURTLEFF, INC., MEDOS S.A. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Required Intervention |