FDA Adverse Event Injury Summary report: N

HAKIM PROGRAMMABLE VALVE W/PRECHAMBER

MDR report key: 2103700 · Received May 23, 2011

Report

Report Number
1226348-2011-00187
Event Type
Injury
Date Received
May 23, 2011
Date of Event
March 28, 2011
Manufacturer
CODMAN & SHURTLEFF, INC., MEDOS S.A.
Product Code
JXG
PMA / PMN Number
K974739
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

HISTORICALLY, FOR COMPLAINTS OF THIS NATURE, EXAMINATIONS OF THE VALVES HAVE REVEALED THE PRESENCE OF BIOLOGICAL DEBRIS WITHIN THE DEVICES WHICH HAVE INTERFERED WITH THEIR PROGRAMMING AND/OR PRESSURE CONTROL MECHANISMS. THE BIOLOGICAL DEBRIS HAS CAUSED THE RETURNED VALVES TO FAIL THE PROGRAMMING AND/OR PRESSURE TESTS AND APPEARS TO HAVE CAUSED THE DIFFICULTIES EXPERIENCED BY THE CUSTOMERS. REVIEW OF THE DEVICE HISTORY RECORDS HAVE CONFIRMED THAT THE VALVES HAVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR IF ANY FURTHER INFO REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CONSIDERED TO BE CLOSED AT THIS TIME.

Description of Event or Problem · 1

AFFILIATE REPORTED THAT THE CUSTOMER COMPLAINED THAT THE VALVE WAS NO LONGER ADJUSTABLE THEREFORE THEY DECIDED TO EXPLANT THE VALVE AND REPLACE IT WITH A NEW ONE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HAKIM PROGRAMMABLE VALVE W/PRECHAMBER SHUNT, CENTRAL NERVOUS SYSTEM AND COMPS JXG CODMAN & SHURTLEFF, INC., MEDOS S.A. NA CLNCOV

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention