FDA Adverse Event Malfunction Summary report: N

GEMINI STONE RETRIEVAL PAIRED WIRE / HELICAL BASKET

MDR report key: 2103580 · Received May 26, 2011

Report

Report Number
3005099803-2011-01793
Event Type
Malfunction
Date Received
May 26, 2011
Date of Event
May 4, 2011
Report Date
May 4, 2011
Manufacturer
BOSTON SCIENTIFIC - SPENCER
Product Code
FFL
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. AT THIS TIME WE ARE UNABLE TO DETERMINE THE RELATIONSHIP BETWEEN THE DEVICE AND THE CAUSE FOR THIS EVENT.

Additional Manufacturer Narrative · 1

IT WAS CONFIRMED THAT THE EVENT OCCURRED PRIOR TO STONE RETRIEVAL. IT IS ALSO NOTED THAT A BASKET NOT OPENING OR CLOSING PRIOR TO STONE RETRIEVAL IS NOT A MEDWATCH REPORTABLE CONDITION.

Description of Event or Problem · 1

PLEASE NOTE: THIS REPORT PERTAINS TO THE SECOND OF FOUR DEVICES. REF: 3005099803-2011-01795, 3005099803-2011-01797, AND 3005099803-2011-01798. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A GEMINI STONE RETRIEVAL BASKET WAS OPENED FOR USE IN A URETEROSCOPY WITH LITHOTRIPSY PROCEDURE. ACCORDING TO THE COMPLAINANT, THE DEVICE WAS TESTED PRIOR TO THE PROCEDURE AND FOUND TO OPEN AND CLOSE WITHOUT PROBLEM. AFTER THE BASKET WAS PUT INTO THE SEMI-RIGID SCOPE IT FAILED TO OPEN AND CLOSE CORRECTLY. THE SCRUB NURSE SAID THAT BASKET HANDLE WAS NOT WORKING CORRECTLY AND BASKET DID NOT OPEN OR CLOSE PROPERLY AFTER GOING THROUGH SCOPE. FOUR GEMINI BASKETS WERE REPORTED TO HAVE THE SAME PROBLEM DURING THE SAME CASE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THE PATIENT IS REPORTED TO BE STABLE.

Description of Event or Problem · 1

PLEASE NOTE: THIS REPORT PERTAINS TO THE SECOND OF FOUR DEVICES. REF: 3005099803-2011-01795, 3005099803-2011-01797, AND 3005099803-2011-01798. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A GEMINI STONE RETRIEVAL BASKET WAS OPENED FOR USE IN A URETEROSCOPY WITH LITHOTRIPSY PROCEDURE. ACCORDING TO THE COMPLAINANT, THE DEVICE WAS TESTED PRIOR TO THE PROCEDURE AND FOUND TO OPEN AND CLOSE WITHOUT PROBLEM. AFTER THE BASKET WAS PUT INTO THE SEMI-RIGID SCOPE IT FAILED TO OPEN AND CLOSE CORRECTLY. THE SCRUB NURSE SAID THAT BASKET HANDLE WAS NOT WORKING CORRECTLY AND BASKET DID NOT OPEN OR CLOSE PROPERLY AFTER GOING THROUGH SCOPE. FOUR GEMINI BASKETS WERE REPORTED TO HAVE THE SAME PROBLEM DURING THE SAME CASE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THE PATIENT IS REPORTED TO BE STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GEMINI STONE RETRIEVAL PAIRED WIRE / HELICAL BASKET DISLODGER, STONE, BASKET, URETERAL, METAL FFL BOSTON SCIENTIFIC - SPENCER M0063303060 13601207

Patients

Seq Age Sex Outcome Treatment
1 60 YR