HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-06540
- Event Type
- Malfunction
- Date Received
- May 26, 2011
- Date of Event
- October 29, 2010
- Report Date
- May 7, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE LOT NUMBER WAS NOT PROVIDED; THEREFORE A BATCH REVIEW CANNOT BE CONDUCTED. THE SAMPLE WAS DISCARDED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MDR WILL BE SENT.
(B)(4). THE SYSTEM ERROR (SE) 2240 WAS IDENTIFIED DURING A REVIEW OF A RETURNED HOMECHOICE (HC) DEVICE; THERE WAS NO REPORT FOR THIS ALARM CALLED IN BY THE CUSTOMER. THE CUSTOMER DISCONTINUED USE OF THE HC MACHINE AND RETURNED IT TO BAXTER. NOT ENOUGH DATA IS AVAILABLE WITHIN THE COMPLAINT TO IDENTIFY ROOT CAUSE; THEREFORE, THE CAUSE WAS NOT DETERMINED. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
A SYSTEM ERROR (SE) 2240 ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THE SYSTEM ERROR OCCURRED ON (B)(6) 2010 . A SE 2240 IS A NON-RECOVERABLE ALARM THAT OCCURS WHEN THE DEVICE HAS DETECTED AIR BEING DRAWN CONTINUOUSLY INTO THE DISPOSABLE. IT IS UNKNOWN IF THIS ALARM OCCURRED DURING PATIENT THERAPY, HOWEVER NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | HOME CHOICE |