FDA Adverse Event Malfunction Summary report: N

HILL-ROM NURSE CALL

MDR report key: 2103549 · Received May 18, 2011

Report

Report Number
2103549
Event Type
Malfunction
Date Received
May 18, 2011
Date of Event
May 13, 2011
Report Date
May 18, 2011
Manufacturer
HILL-ROM
Product Code
ILQ
Product Problem
Yes
Report Source
User Facility report
Reporter Location
IA, US

Narratives

Description of Event or Problem · 1

INITIALLY THERE WAS NO AUDIO TO THE REHAB DINING ROOM STAFF STATION. WHEN BIOMEDICAL STAFF TRIED TO RESET IT, IT WOULD NOT RESET. DURING A CONFIGURATION EXAMINATION, IT WAS DETERMINED THE DINING ROOM WAS IN THE CONFIGURATION, BUT NOT IN THE RESET LIST. CLIENT PC WAS DEFECTIVE AND REP ORDERED A NEW ONE AND WAS INSTALLED SAME DAY.======================MANUFACTURER RESPONSE FOR NURSE CALL, HILL-ROM NURSE CALL (PER SITE REPORTER)======================REP DETERMINED THE CLIENT PC WAS DEFECTIVE. PART WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HILL-ROM NURSE CALL NURSE CALL ILQ HILL-ROM NCM NURSE COMMUNICATION MODULE *

Patients

Seq Age Sex Outcome Treatment
1 *