FDA Adverse Event Injury Summary report: N

ULTRAFLEX

MDR report key: 2103529 · Received May 26, 2011

Report

Report Number
3005099803-2011-01816
Event Type
Injury
Date Received
May 26, 2011
Date of Event
May 4, 2011
Report Date
May 4, 2011
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
JCT
PMA / PMN Number
K963241
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINANT WAS UNABLE TO REPORT THE PART OR LOT NUMBER; THEREFORE THE MANUFACTURE DATE AND EXPIRATION DATE ARE UNKNOWN. HOWEVER THE COMPLAINANT STATED THAT THE DEVICE WAS USED PRIOR TO THE EXPIRATION DATE THE COMPLAINANT INDICATED THAT THE DEVICE REMAINS IMPLANTED AND WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. IF ANY FURTHER RELEVANT INFORMATION IS IDENTIFIED, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ULTRAFLEX TRACHEOBRONCHIAL STENT WAS IMPLANTED WITHIN THE TRACHEA OF A PATIENT IN (B)(6) 1999. ACCORDING TO THE COMPLAINANT, THE PATIENT PRESENTED WITH A TRACHEAL LESION AS A RESULT OF TRACHEAL MALACIA. IT IS UNKNOWN IF THE TRACHEAL MALACIA WAS CAUSED BY CANCER. THE EXACT SIZE OF THE LESION IS UNKNOWN. ON (B)(6), 2011, THE PATIENT WAS HAVING TROUBLE BREATHING, SO THE PHYSICIAN PERFORMED AN ENDOSCOPIC PROCEDURE WHERE IT WAS DISCOVERED THAT ONE STENT WIRE HAD UNRAVELED AND WAS PROTRUDING INTO THE PATIENT'S TRACHEAL TISSUE. THE STENT WIRE DID NOT CAUSE ANY TISSUE DAMAGE, IT JUST CAUSED AN INCREASE OF MUCOUS TO BUILD UP. THE STENT WAS NOT BLOCKED WITH ANY TISSUE; JUST A MUCOUS PLUG. ON (B)(6), 2011 THE PHYSICIAN SUCCESSFULLY CUT THE STENT WIRE USING ENDOSCOPIC SCISSORS, AND REMOVED IT WITHOUT DIFFICULTY. THE STENT IS STILL IMPLANTED, AND NO FURTHER INTERVENTION IS PLANNED. THE PATIENT IS IN GOOD CONDITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRAFLEX PROSTHESIS, TRACHEAL, EXPANDABLE JCT BOSTON SCIENTIFIC - GALWAY UNK449

Patients

Seq Age Sex Outcome Treatment
1 72 YR Required Intervention