MINICAP TRANSFER SET
Report
- Report Number
- 1423500-2011-06550
- Event Type
- Injury
- Date Received
- May 26, 2011
- Date of Event
- February 1, 2011
- Report Date
- May 6, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- KDJ
- PMA / PMN Number
- K882498
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS IS REPORT 4 OF 4 INVOLVED IN THIS PERITONITIS EVENT. THE SAMPLE IS NOT AVAILABLE. A FOLLOW-UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). THE ROOT CAUSE OF THE REPORTED CONDITION OF PERITONITIS WAS USE ERROR- POOR ASEPTIC TECHNIQUE. A BATCH REVIEW WAS PERFORMED FOR THE POTENTIALLY ASSOCIATED LOT NUMBERS (H10J11054). THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE. A LABELING REVIEW WAS PERFORMED AND THE LABELING WAS FOUND TO PROVIDE AMPLE INSTRUCTIONS RELATED TO PREVENTION OF THE USE ERROR IN ASEPTIC TECHNIQUE. BAXTER HAS RECEIVED SIMILAR REPORTS FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.
THIS REPORT WAS RECEIVED FROM GLOBAL PHARMACOVIGILANCE. THIS IS A SPONTANEOUS REPORT BY A CONSUMER WITH SUPPLEMENTAL INFORMATION BY A NURSE FROM THE USA OF PATIENT MADE MISTAKE/TOUCH CONTAMINATION AND PERITONITIS WITH CULTURE POSITIVE FOR STAPH EPIDERMIDIS COAGULASE NEG STAPH IN A (B)(6) FEMALE PATIENT COINCIDENT WITH DIANEAL PD4 AMBUFLEX THERAPY. ON AN UNREPORTED DATE, THE PATIENT BEGAN TREATMENT WITH DIANEAL PD4 AMBUFLEX THERAPY (LOT NUMBER, DOSE, AND FREQUENCY NOT REPORTED) INTRAPERITONEALLY FOR PERITONEAL DIALYSIS (PD). DURING A CALL WITH BAXTER CUSTOMER SERVICES, THE FOLLOWING WAS REPORTED BY THE CONSUMER. ON AN UNREPORTED DATE IN 2011, THE PATIENT MADE A MISTAKE/TOUCH CONTAMINATION. ON (B)(6) 2011, THE PATIENT EXPERIENCED PERITONITIS. THE PATIENT WAS HOSPITALIZED FROM (B)(6) 2011. TREATMENT WAS NOT REPORTED. THE PATIENT WAS RECOVERING FROM THE PERITONITIS AND THE OUTCOME FOR THE MADE MISTAKE/TOUCH CONTAMINATION WAS NOT REPORTED. DIANEAL THERAPY WAS ONGOING. AN OPINION OF CAUSALITY WAS NOT REPORTED. THE NURSE STATED THE PERITONITIS WAS NOT RELATED TO DIANEAL THERAPY AND WAS DUE TO THE PATIENT MADE MISTAKE/TOUCH CONTAMINATION. AN OPINION OF CAUSALITY WAS NOT REPORTED FOR THE PATIENT MADE MISTAKE/TOUCH CONTAMINATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MINICAP TRANSFER SET | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR | Hospitalization | DIANEAL PD4 AMBUFLEX| HOMECHOICE DEVICE |