FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2103510 · Received May 26, 2011

Report

Report Number
2531779-2011-03720
Event Type
Malfunction
Date Received
May 26, 2011
Date of Event
May 2, 2011
Report Date
May 2, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(6). THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS WERE INTERMITTENTLY ACTIVATING PUMP FUNCTIONS WHEN PRESSED. ADHESIVE WAS FOUND UNDER THE KEYPAD BUTTON CONTACTS.

Description of Event or Problem · 1

THE DISTRIBUTOR REPORTED THAT THE UP AND DOWN BUTTONS DID NOT ACTIVATE DESIRED PUMP FUNCTIONS WHEN PRESSED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 1200 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1