FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2103510
·
Received May 26, 2011
Report
- Report Number
- 2531779-2011-03720
- Event Type
- Malfunction
- Date Received
- May 26, 2011
- Date of Event
- May 2, 2011
- Report Date
- May 2, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
(B)(6). THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS WERE INTERMITTENTLY ACTIVATING PUMP FUNCTIONS WHEN PRESSED. ADHESIVE WAS FOUND UNDER THE KEYPAD BUTTON CONTACTS.
Description of Event or Problem · 1
THE DISTRIBUTOR REPORTED THAT THE UP AND DOWN BUTTONS DID NOT ACTIVATE DESIRED PUMP FUNCTIONS WHEN PRESSED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 1200 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |