FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2103478 · Received May 26, 2011

Report

Report Number
1423500-2011-06531
Event Type
Malfunction
Date Received
May 26, 2011
Date of Event
April 21, 2011
Report Date
May 11, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED AND EVALUATED. ADDITIONAL ISSUE OF IIPV WAS CONFIRMED IN THE THERAPY LOG, BUT NOT DUPLICATED DURING PAL EVALUATION. THE ASSIGNABLE CAUSE WAS: INSUFFICIENT DRAIN. USE ERROR. TIDAL UF REMOVAL SET TOO LOW. A REVIEW OF THE PREVIOUS SERVICE RECORD, (B)(6) 2009, SHOWS THE DEVICE PASSED ALL REQUIRED TESTS AND CALIBRATIONS PRIOR TO ITS RELEASE FROM THE (B)(4) FACILITY. NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE ISSUE OF IIPV. THE PREVIOUS RETURN OF THE DEVICE WAS NOT FOR ANY ISSUE RELATED TO. DEVICE MET SPECIFICATIONS RELATIVE TO THE IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THIS ISSUE HAS BEEN ESCALATED TO CAPA.

Description of Event or Problem · 1

DURING EVALUATION, AN ADDITIONAL ISSUE OF AN INCREASED INTRAPERITONEAL VOLUME (IIPV) EVENT WAS FOUND IN THE PATIENT EVENT LOGS: OCCURRENCE DATE (B)(6) 2011 WITH DRAIN VOLUME OF 3239 MILLILITERS (ML) DURING CYCLE 4. THIS DRAIN VOLUME MEETS BAXTER'S IIPV CRITERIA. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 62 YR HOME CHOICE CASSETTE