FDA Adverse Event Injury Summary report: N

STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL

MDR report key: 2103432 · Received May 26, 2011

Report

Report Number
3005099803-2011-01714
Event Type
Injury
Date Received
May 26, 2011
Date of Event
May 4, 2011
Report Date
May 4, 2011
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
FGO
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GB
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINANT INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION BECAUSE IT WAS DISPOSED; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. AT THIS TIME, WE ARE UNABLE TO DETERMINE THE RELATIONSHIP BETWEEN THIS DEVICE AND THE CAUSE FOR THIS EVENT. IF THERE IS ANY FURTHER RELEVANT INFORMATION A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Additional Manufacturer Narrative · 1

FOLLOW UP INFORMATION RECEIVED ON (B)(6), 2011 STATING STONE CONE NITINOL RETRIEVAL BASKET TIP WAS REMOVED FROM THE PATIENT VIA A PERCUTANEOUS NEPHROLITHOTOMY PROCEDURE ON (B)(6), 2011. THE PATIENT WAS REPORTED TO BE "FIT AND WELL". THE FACILITY REPORTED THAT THE TIP FRAGMENT WILL BE RETURNED UPON COMPLETION OF HOSPITAL REPORT.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A STONE CONE RETRIEVAL DEVICE WAS USED IN A FLEXIBLE UTEROSCOPY PROCEDURE (PATIENT IDENTIFIER, AGE, SEX, AND WEIGHT UNKNOWN). ACCORDING TO THE COMPLAINANT, THE PHYSICIAN DEPLOYED THE STONE CONE BEHIND THE STONE AND WAS LASERING THE UPPER STONE WHEN A SECTION OF THE STONE CONE SPIRAL TIP BROKE OFF AND FELL INTO THE RENAL PELVIS. PHYSICIAN WAS UNABLE TO RETRIEVE THE BROKEN SECTION WITH GRASPING FORCEPS. ADDITIONAL SURGERY MAY BE REQUIRED DEPENDING ON THE RESULTS OF A CT SCAN. AT THIS TIME THERE IS NO FURTHER INFORMATION AVAILABLE. THE PATIENT IS REPORTED TO BE STABLE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A STONE CONE RETRIEVAL DEVICE WAS USED IN A FLEXIBLE UTEROSCOPY PROCEDURE (PATIENT IDENTIFIER, AGE, SEX, AND WEIGHT UNKNOWN). ACCORDING TO THE COMPLAINANT, THE PHYSICIAN DEPLOYED THE STONE CONE BEHIND THE STONE AND WAS LASERING THE UPPER STONE WHEN A SECTION OF THE STONE CONE SPIRAL TIP BROKE OFF AND FELL INTO THE RENAL PELVIS. PHYSICIAN WAS UNABLE TO RETRIEVE THE BROKEN SECTION WITH GRASPING FORCEPS. ADDITIONAL SURGERY MAY BE REQUIRED DEPENDING ON THE RESULTS OF A CT SCAN. AT THIS TIME THERE IS NO FURTHER INFORMATION AVAILABLE. THE PATIENT IS REPORTED TO BE STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STONE CONE NITINOL UROLOGICAL RETRIEVAL COIL DISLODGER, STONE, FLEXIBLE FGO BOSTON SCIENTIFIC - MARLBOROUGH M0063903100 0000709961

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention