FDA Adverse Event Malfunction Summary report: N

SWARTZ 8.5F BRAIDED TRANS INTRODUCER, SL0, 63CM

MDR report key: 2103368 · Received May 17, 2011

Report

Report Number
3005188751-2011-00054
Event Type
Malfunction
Date Received
May 17, 2011
Date of Event
April 29, 2011
Report Date
May 13, 2011
Manufacturer
ST. JUDE MEDICAL, AF DIVISION
Product Code
DYB
PMA / PMN Number
K052644
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MANUFACTURER - ONE 8.5F BRAIDED SWARTZ INTRODUCER WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION REVEALED DAMAGE TO THE WHITE HANDLE ON THE STOPCOCK AT THE STOPPER, WHICH HAD BEEN OVER ROTATED CAUSING THE HANDLE TO GO BEYOND THE STOP POINT AND COMPLETELY ROTATE WHEN TURNED. A SCREW ON SYRINGE WAS FILLED WITH GREEN DYE AND WAS ATTACHED TO ONE PORT OF THE STOPCOCK AND THE CATHETER SHAFT WAS CLAMPED OFF. THE GREEN DYE WAS PUSHED THROUGH THE STOPCOCK WITH THE HANDLE BEING ROTATED; NO LEAK WAS OBSERVED BETWEEN THE HANDLE WALL AND THE STOPCOCK. IF THE STOPCOCK WAS OVER ROTATED AND CAME IN CONTACT WITH THE UNUSED PORT, THEN GREEN DYE WOULD EXIT THROUGH THE UNUSED STOPCOCK PORT. REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THIS LOT MET MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT.

Description of Event or Problem · 1

THIS COMPLAINT IS REPORTABLE BASED ON ANALYSIS COMPLETED ON (B)(6) 2011. IT WAS REPORTED THAT THE LUER HUB ON THE STOPCOCK OF THE SWARTZ INTRODUCER WOULD NOT LOCK ONTO THE LUER TIP OF THE NON SJM INFUSION TUBING. THE PHYSICIAN NOTED THAT THE CONNECTION WAS LOOSE AS HE WAS ATTACHING THE PRESSURE TUBING WHILE THE INTRODUCER SHEATH WAS IN THE PATIENT. THE TUBING SET WAS EXCHANGED AND THE SAME ISSUE WAS NOTED. THE INTRODUCER SHEATH WAS THEN EXCHANGED AND THE PROCEDURE WAS COMPLETED WITH NO CONSEQUENCES TO THE PATIENT. DEVICE INVESTIGATION REVEALED DAMAGE TO THE STOPCOCK CAUSED BY OVER-ROTATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SWARTZ 8.5F BRAIDED TRANS INTRODUCER, SL0, 63CM SWARTZ, 8.5F, INTRO, SL0 DYB ST. JUDE MEDICAL, AF DIVISION 407451 3310506

Patients

Seq Age Sex Outcome Treatment
1