FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2103357
·
Received May 17, 2011
Report
- Report Number
- 1720753-2011-07379
- Event Type
- Malfunction
- Date Received
- May 17, 2011
- Date of Event
- April 27, 2011
- Report Date
- May 17, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REP RESEATED THE MAINFRAME PRINTED CIRCUIT BOARDS AND TIGHTENED THE HIGH VOLTAGE CONNECTIONS AND PERFORMED A BATTERY LOAD TEST. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A CHARGER FAILED ERROR MESSAGE DURING FLUOROSCOPIC X-RAY. NO PT INJURY WAS REPORTED.