FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2103357 · Received May 17, 2011

Report

Report Number
1720753-2011-07379
Event Type
Malfunction
Date Received
May 17, 2011
Date of Event
April 27, 2011
Report Date
May 17, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REP RESEATED THE MAINFRAME PRINTED CIRCUIT BOARDS AND TIGHTENED THE HIGH VOLTAGE CONNECTIONS AND PERFORMED A BATTERY LOAD TEST. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A CHARGER FAILED ERROR MESSAGE DURING FLUOROSCOPIC X-RAY. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1