FDA Adverse Event
Malfunction
Summary report: N
6800
MDR report key: 2103346
·
Received May 17, 2011
Report
- Report Number
- 1720753-2011-07370
- Event Type
- Malfunction
- Date Received
- May 17, 2011
- Date of Event
- April 26, 2011
- Report Date
- May 17, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REPRESENTATIVE REPLACED THE SINGLE BOARD COMPUTER AND THE HARD DISK DRIVE, LOADED THE SOFTWARE AND THE CONFIGURATION FILES AS WELL AS INSTALLED THE WORKSTATION REAR COVER FASTENER KIT AND THE ETHERNET CABLE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 6800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 6800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |