FDA Adverse Event
Malfunction
Summary report: N
STENOSCOP
MDR report key: 2103317
·
Received May 17, 2011
Report
- Report Number
- 9617766-2011-01066
- Event Type
- Malfunction
- Date Received
- May 17, 2011
- Date of Event
- April 26, 2011
- Report Date
- May 17, 2011
- Manufacturer
- GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD (GE PRIVATE)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE LIMIT SWITCH WAS REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM'S COLLIMATOR SHUTTERS CLOSED AUTOMATICALLY WITHOUT COMMAND WHEN THE MAGNIFICATION MODE WAS SELECTED. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STENOSCOP | FLUOROSCOPIC X-RAY | JAA | GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD (GE PRIVATE) | STENOSCOP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |