SUTURE ANCHOR, BIO- PUSHLOCK, 3.5MM
Report
- Report Number
- 1220246-2011-00087
- Event Type
- Injury
- Date Received
- May 26, 2011
- Date of Event
- March 28, 2011
- Report Date
- May 6, 2011
- Manufacturer
- ARTHREX, INC.
- Product Code
- HWC
- PMA / PMN Number
- K061863
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MT, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
NO FURTHER PATIENT INFORMATION WAS PROVIDED AT THE TIME OF THIS REPORT OR MADE AVAILABLE IN RESPONSE TO FOLLOW-UP COMMUNICATION. NO ADDITIONAL ADVERSE CONSEQUENCES HAVE BEEN REPORTED FROM THIS EVENT. THIS DEVICE IS USED FOR TREATMENT. THE DEVICE WAS REQUESTED FOR EVALUATION BUT WAS NOT RETURNED, THEREFORE THE COMPLAINANT'S EVENT COULD NOT BE VERIFIED. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED FROM THE INFORMATION AVAILABLE AND WITHOUT DEVICE EVALUATION. DEVICE HISTORY RECORD REVEALED NOTHING RELEVANT TO THIS EVENT; NO ISSUES WERE FOUND WITH THE STERILIZATION OF THIS LOT. THIS DEVICE IS SUPPLIED STERILE. BASED ON THE INFORMATION PROVIDED, A DEFINITIVE CAUSE FOR THE POST-OPERATIVE INFECTION COULD NOT BE DETERMINED. THE MOST LIKELY CAUSE FOR THIS TYPE OF EVENT IS A NOSOCOMIAL SOURCE OR PATIENT NON-COMPLIANCE WITH POST-OP WOUND CARE DIRECTIONS. THIS IS THE FIRST COMPLAINT OF THIS TYPE FOR THIS PART/LOT COMBINATION. THE POTENTIAL CAUSES OF THIS EVENT ARE BEING COMMUNICATED TO THE EVENT REPORTER. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED. DISCARDED BY FACILITY.
SIGNIFICANT INFECTION AFTER THREE WEEKS POST OP. DOS WAS (B)(6) 2011. ROTATOR CUFF REPAIR. (B)(6) POST-OP, REPORTED PAIN AT INSERTION SITE, INFECTION. FIRST SEEN FOR FLUSHING OF INFECTED WOUND ON (B)(6) 2011. MAY BE SEEN LATER IN WEEK FOR FOLLOW UP/2ND FLUSHING OF WOUND. REMOVED IMPLANTS AND ALL SUTURE TODAY. PATIENT`S CUFF IS HEALED. NO ADDITIONAL IMPLANTS WERE NEEDED .PATIENT TREATED WITH ANTIBIOTIC, NO CANNULA WAS USED .BLOOD AND WOUND CULTURES ARE NEGATIVE. PATIENT IS DOING FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SUTURE ANCHOR, BIO- PUSHLOCK, 3.5MM | SCREW, FIXATION, BONE | HWC | ARTHREX, INC. | 265337 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Other | (B)(4), LOT 389622 |