ACL TIGHTROPE
Report
- Report Number
- 1220246-2011-00083
- Event Type
- Injury
- Date Received
- May 26, 2011
- Date of Event
- April 15, 2011
- Report Date
- May 3, 2011
- Manufacturer
- ARTHREX, INC.
- Product Code
- HTY
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
PATIENT DEMOGRAPHICS (AGE AT TIME OF EVENT, DATE OF BIRTH, GENDER, WEIGHT) WERE REQUESTED BUT NOT PROVIDED. NO FURTHER PATIENT INFORMATION WAS PROVIDED AT THE TIME OF THIS REPORT OR MADE AVAILABLE IN RESPONSE TO FOLLOW-UP COMMUNICATION. NO ADDITIONAL ADVERSE CONSEQUENCES HAVE BEEN REPORTED FROM THIS EVENT. THIS DEVICE IS USED FOR TREATMENT. THE DEVICE WAS RECEIVED AND AN EVALUATION WAS CONDUCTED. DEVICE HISTORY RECORD NOT PERFORMED AS THE LOT NUMBER WAS NOT PROVIDED. COMPLAINT CONFIRMED. VISUAL INSPECTION REVEALED THAT THE SUTURE LOOP APPEARS TO HAVE BEEN CUT AS WELL AS HAVING NICKS AND GOUGES ON THE LOOP. THE MOST LIKELY CAUSE(S) FOR TORN OR BROKEN SUTURES INCLUDE NICKING THE SUTURE WITH ANOTHER INSTRUMENT AND/OR FRAYING FROM SHARP EDGES OF THE BONE TUNNEL. THIS IS THE FIRST COMPLAINT OF THIS TYPE FOR THIS PART/LOT COMBINATION. THE POTENTIAL CAUSES OF THIS EVENT ARE BEING COMMUNICATED TO THE EVENT REPORTER. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED.
PULLED ACL TIGHTROPE UP INTO FEMORAL SOCKET, BUT PULLED TOO HARD AND THE BUTTON CAME OUT OF THE SKIN. THE SURGEON THEN MADE A SMALL INCISION ON THE OUTSIDE OF THE FEMUR (SKIN), AND TURNED THE BUTTON VERTICAL TO PUSH BACK DOWN TO BONE. THEN PULLED BACK ON THE GRAFT TO CONFIRM IT WAS SEATED AND THE GRAFT AND ACL TIGHTROPE CAME BACK OUT AND SAW THAT THE LOOP HAD BROKE. EVERYTHING WAS REMOVED AND STARTED OVER AGAIN WITH A NEW (B)(4), AND COMPLETED CASE WITH NO PROBLEMS
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACL TIGHTROPE | PIN, FIXATION, SMOOTH | HTY | ARTHREX, INC. | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |