NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2011-00312
- Event Type
- Malfunction
- Date Received
- May 25, 2011
- Date of Event
- April 28, 2011
- Report Date
- May 3, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE COMPLAINT RD900AEU NEOPUFF INFANT RESUSCITATOR IS EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE RECEIVE THE COMPLAINT DEVICE AND HAVE COMPLETED OUR INVESTIGATION.
(B)(4). METHOD: THE FASCIA AND THE VALVE SYSTEM OF THE COMPLAINT RD900AEU NEOPUFF INFANT RESUSCITATOR WERE RETURNED TO FPH (B)(4) FOR INSPECTION. THE VALVE SYSTEM WAS FITTED WITH A TEST MANOMETER AND PERFORMANCE TESTED BASED ON THE NEOPUFF TECHNICAL MANUAL. THE PIP VALVE WAS CHECKED FOR VIBRATION BY APPLYING DIFFERENT GAS FLOW RATES. RESULTS: THE RETURNED VALVE SYSTEM PASSED THE PERFORMANCE TEST SPECIFIED IN THE NEOPUFF TECHNICAL MANUAL; HOWEVER, THE PIP VALVE FAILED THE VIBRATION TEST. IT WAS NOTED TO BE NOISY AT CERTAIN PRESSURES AND AT DIFFERENT GAS FLOW RATES, CONFIRMING THE REPORTED FAULT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 110223. CONCLUSION: THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. WE WERE UNABLE TO DETERMINE WHAT CAUSED THE FAULT OBSERVED BY THE HOSPITAL. ALL NEOPUFF UNITS ARE PERFORMANCE TESTED, INCLUDING VALVE VIBRATION MONITORING, PRIOR TO LEAVING THE PRODUCTION LINE AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THE VALVE BECAME DEFECTIVE POST PRODUCTION. IT IS MOST LIKELY THAT THE NEOPUFF HAS BEEN SUBJECTED TO IMPACT DURING TRANSPORTATION WITHOUT INCURRING ANY PHYSICAL DAMAGE, WHICH COULD HAVE CONTRIBUTED TO THE REPORTED FAULT. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: "DROPPING THE NEOPUFF / PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT". A REPLACEMENT VALVE ASSEMBLY KIT HAS BEEN SUPPLIED TO THE HOSPITAL.
A HOSPITAL IN (B)(6) REPORTED THAT THE VALVE ASSEMBLY OF AN RD900AEU NEOPUFF INFANT RESUSCITATOR MADE A LOUD VIBRATING NOISE WHEN THE PEAK INSPIRATORY PRESSURE (PIP) VALVE WAS ADJUSTED. THIS WAS NOTICED PRIOR TO PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.
A HOSPITAL IN (B)(6) REPORTED THAT THE VALVE ASSEMBLY OF AN RD900AEU NEOPUFF INFANT RESUSCITATOR MADE A LOUD VIBRATING NOISE WHEN THE PEAK INSPIRATORY PRESSURE (PIP) WAS ADJUSTED. THIS WAS NOTICED PRIOR TO PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | FISHER & PAYKEL HEALTHCARE LIMITED | RD900AEU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |