NAVI-STAR¿ THERMO-COOL¿ ELECTROPHYSIOLOGY CATHETER
Report
- Report Number
- 9673241-2011-00055
- Event Type
- Malfunction
- Date Received
- May 25, 2011
- Date of Event
- April 27, 2011
- Report Date
- April 28, 2011
- Manufacturer
- BIOSENSE WEBSTER, INC. (JUAREZ)
- Product Code
- OAD
- PMA / PMN Number
- P030031
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4). THE EVENT IN THIS COMPLAINT WAS RELATED TO A WHITISH SUBSTANCE RESEMBLING FIBRIN NOTED ON THE CATHETER TIP AFTER A RIGHT-SIDED ATRIAL FLUTTER PROCEDURE. UPON RECEIPT OF THE RETURNED DEVICE, VISUAL INSPECTION EXHIBITED FOREIGN MATERIAL ON THE CATHETER TIP. THE DEVICE WAS TESTED FOR ELECTRICAL PERFORMANCE, STOCKERT COMPATIBILITY, THERMAL SENSOR RESPONSE AND PATENCY FLOW TESTS. THE ELECTRICAL RESISTANCE AND LEAKAGE TESTS WERE FOUND WITHIN SPECIFICATIONS. THE THERMOCOUPLE SENSOR PASSED WHEN IMMERSED IN FLUID WITH CONTROLLED TEMPERATURE. THE CATHETER INTERFACED WITH THE STOCKERT OBTAINING ACCEPTABLE ABLATION READINGS. HOWEVER, THE PATENCY FLOW TEST FAILED. THE CATHETER WAS NOT IRRIGATING WHEN TESTED. FURTHER INVESTIGATION REVEALED THAT THE IRRIGATION TUBING WAS OBSTRUCTED WITH SOME BROWNISH MATERIAL NOT ALLOWING FLUID TO FLOW THROUGH. THE BROWNISH RESIDUE/ MATERIAL WAS NOT PART OF THE COMPLAINT INITIALLY REPORTED. BOTH THE WHITISH AND BROWNISH MATERIALS FOUND WERE ANALYZED USING THE FT-IR METHOD FOR MATERIAL IDENTIFICATION. THE FT-IR RESULTS SHOWED THAT BOTH MATERIALS CONSISTED OF PROTEIN THAT CAN BE FOUND IN THE AORTIC AND/OR VEIN TISSUES. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, DHR REVIEW VERIFIED THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE TO DOCUMENTED SPECIFICATION AND PROCEDURES. COMPLAINT CONDITION WAS CONFIRMED.
AFTER MULTIPLE ATTEMPTS TO RETRIEVE THE DEVICE, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. PRODUCT HAS NOT BEEN RETURNED FOR INVESTIGATION AS INITIALLY REPORTED.(B)(4).
IT WAS REPORTED THAT AFTER A RIGHT-SIDED ATRIAL FLUTTER PROCEDURE, UPON RETRIEVAL, THE CATHETER APPEARED TO BE WRAPPED BY A WHITISH SUBSTANCE RESEMBLING FIBRIN. THE SURGEON TRIED TO REMOVE THE SUBSTANCE, HOWEVER IT REMAINED STUCK ON THE CATHETER TIP. IT WAS MENTIONED THAT THERE WAS NO SHARP EDGES ON THE CATHETER TIP. THE PHYSICIAN ENCOUNTERED SOME DIFFICULTIES IN ABLATING AND BLOCKING THE RIGHT ISTHMUS. HE ALSO HAD OTHER UNSPECIFIED DIFFICULTIES AFTER HAVING CHANGED THE CATHETER. IT WAS STATED THAT THIS MAY BE DUE TO UNCOMMON PATIENT ANATOMY. THERE WAS NO PATIENT INJURY REPORTED. THERE WAS NO MEDICAL INTERVENTION PERFORMED. THE PROCEDURE WAS COMPLETED BY USING A NEW CATHETER WITHOUT ANY PROBLEM. THE USERS WERE UNSURE IF THEY CHECKED THE CATHETER TIP BEFORE INSERTING IT TO THE PATIENT AS RECOMMENDED IN THE IFU.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NAVI-STAR¿ THERMO-COOL¿ ELECTROPHYSIOLOGY CATHETER | CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER | OAD | BIOSENSE WEBSTER, INC. (JUAREZ) | D-1197-17-S | 15302545M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |