FDA Adverse Event
Injury
Summary report: N
SPECTRA WAVEWRITER
MDR report key: 21031365
·
Received December 27, 2024
Report
- Report Number
- 3006630150-2024-09104
- Event Type
- Injury
- Date Received
- December 27, 2024
- Date of Event
- October 3, 2024
- Report Date
- December 27, 2024
- Manufacturer
- BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
- Product Code
- LGW
- UDI-DI
- 08714729951254
- PMA / PMN Number
- P030017
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2317700. MODEL: SC-2317-70. SERIAL: (B)(6). BATCH: 5122936/5119180.
Description of Event or Problem · 0
IT WAS REPORTED THAT THE PATIENT WAS FREQUENTLY CHARGING THE IMPLANTABLE PULSE GENERATOR (IPG). IT WAS ALSO STATED THAT ONE OF THE PATIENT'S LEADS HAD HIGH IMPEDANCE. THAT PATIENT UNDERWENT A PROCEDURE WHEREIN THE IPG AND LEADS WERE REPLACED. THE PATIENT WAS DOING WELL POSTOPERATIVELY AND THE EXPLANTED DEVICES WILL NOT BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1718410 | SPECTRA WAVEWRITER | STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF | LGW | BOSTON SCIENTIFIC NEUROMODULATION CORPORATION | SC-1160 | 348910 | 08714729951254 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Female | Required Intervention |