FDA Adverse Event Injury Summary report: N

SPECTRA WAVEWRITER

MDR report key: 21031365 · Received December 27, 2024

Report

Report Number
3006630150-2024-09104
Event Type
Injury
Date Received
December 27, 2024
Date of Event
October 3, 2024
Report Date
December 27, 2024
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
Product Code
LGW
UDI-DI
08714729951254
PMA / PMN Number
P030017
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2317700. MODEL: SC-2317-70. SERIAL: (B)(6). BATCH: 5122936/5119180.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS FREQUENTLY CHARGING THE IMPLANTABLE PULSE GENERATOR (IPG). IT WAS ALSO STATED THAT ONE OF THE PATIENT'S LEADS HAD HIGH IMPEDANCE. THAT PATIENT UNDERWENT A PROCEDURE WHEREIN THE IPG AND LEADS WERE REPLACED. THE PATIENT WAS DOING WELL POSTOPERATIVELY AND THE EXPLANTED DEVICES WILL NOT BE RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1718410 SPECTRA WAVEWRITER STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW BOSTON SCIENTIFIC NEUROMODULATION CORPORATION SC-1160 348910 08714729951254

Patients

Seq Age Sex Outcome Treatment
1 55 YR Female Required Intervention