FDA Adverse Event Malfunction Summary report: N

GYNECARE X-TRACT MORCELLATOR

MDR report key: 2103101 · Received May 25, 2011

Report

Report Number
2210968-2011-00671
Event Type
Malfunction
Date Received
May 25, 2011
Date of Event
January 20, 2011
Report Date
May 2, 2011
Manufacturer
ETHICON
Product Code
HET
PMA / PMN Number
K993801
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) - DAMAGE TO DRIVE CONNECTOR OR COUPLING. CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE REPORTED SYMPTOM WAS DUPLICATED. THE CAUSE WAS DUE TO A DAMAGED MOTOR COUPLER. IN ADDITION, A REVIEW OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED AND THE DEVICE MET ALL FINISHED GOODS RELEASE CRITERIA.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A GYNECOLOGICAL PROCEDURE ON (B)(6) 2011. DURING THE PROCEDURE, THE DEVICE WAS NOT WORKING SMOOTHLY AND MADE A LOUD GRINDING NOISE. A SECOND DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GYNECARE X-TRACT MORCELLATOR LAPAROSCOPIC MORCELLATOR HET ETHICON NA NA

Patients

Seq Age Sex Outcome Treatment
1