FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2103027 · Received May 25, 2011

Report

Report Number
2531779-2011-03700
Event Type
Malfunction
Date Received
May 25, 2011
Date of Event
February 25, 2011
Report Date
April 26, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PUMP WAS RETURNED TO ANIMAS. ON (B)(6) 2011 EVALUATION REVEALED A BROKEN SOLDER TRACE ON A FORCE SENSOR PIN. REVIEW OF THE PUMP DOWNLOAD HISTORY REVEALED LOSS OF PRIME ALARMS ASSOCIATED WITH ZERO FORCE. THE PUMP WAS EXERCISED WITH NO LOSS OF PRIME WARNINGS DUPLICATED. THE PUMP WAS REWOUND, LOADED, PRIMED AND EXERCISED WITH NO ALARMS OCCURRING. THE FORCE SENSOR WAS OUT OF CALIBRATION.

Description of Event or Problem · 1

EVALUATION REVEALED A BROKEN SOLDER TRACE ON A FORCE SENSOR PIN. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 52 YR