FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2102981 · Received May 25, 2011

Report

Report Number
2531779-2011-03697
Event Type
Malfunction
Date Received
May 25, 2011
Date of Event
February 21, 2011
Report Date
April 26, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CORRECTION # = 2531779-03/24/2010/003-R. ON (B)(4), 2011 EVALUATION REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. REVIEW OF THE PUMP HISTORY REVEALED LOSS OF PRIME ALARMS ASSOCIATED WITH LOW FORCE. REWIND, LOAD AND PRIME STEPS WERE PERFORMED SUCCESSFULLY WITH NO ALARMS. THE PUMP WAS EXERCISED WITH NO ALARMS OCCURRING.

Description of Event or Problem · 1

EVALUATION REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 62 YR