FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2102981
·
Received May 25, 2011
Report
- Report Number
- 2531779-2011-03697
- Event Type
- Malfunction
- Date Received
- May 25, 2011
- Date of Event
- February 21, 2011
- Report Date
- April 26, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CORRECTION # = 2531779-03/24/2010/003-R. ON (B)(4), 2011 EVALUATION REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. REVIEW OF THE PUMP HISTORY REVEALED LOSS OF PRIME ALARMS ASSOCIATED WITH LOW FORCE. REWIND, LOAD AND PRIME STEPS WERE PERFORMED SUCCESSFULLY WITH NO ALARMS. THE PUMP WAS EXERCISED WITH NO ALARMS OCCURRING.
Description of Event or Problem · 1
EVALUATION REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |