AIR DERMATOME HANDPIECE
Report
- Report Number
- 1526350-2011-00095
- Event Type
- Malfunction
- Date Received
- May 12, 2011
- Date of Event
- April 1, 2011
- Report Date
- April 14, 2011
- Manufacturer
- ZIMMER SURGICAL
- Product Code
- GFD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- UNKNOWN
Narratives
THE DEVICE WAS RETURNED TO THE MFR FOR REPAIR. THE SERVICE RECORDS INDICATE THAT THE DEVICE IS 7 YEARS OLD THE LAST REPAIR WAS ON 07/15/2009, FOR A NON RELATED ISSUE. THE INSPECTION FOUND THAT THE DEVICE RAN BELOW SPECIFICATION AND THE THICKNESS LEVEL SETTINGS WERE OUT OF SPECIFICATION AT 0 AND 30 POSITIONS. THE LIKELY CAUSE OF THE REPORTED COMPLAINT WAS AN ERRATIC MOTOR AND THE DEVICE OUT OF CALIBRATION. THESE ARE DUE TO A LACK OF PREVENTATIVE MAINTENANCE. THIS IS A RE-USABLE DEVICE, SUBJECT TO NORMAL WEAR AND TEAR. THE ZIMMER AIR DERMATOME SHOULD BE RETURNED EVERY 12 MONTHS AND THE HOSE EVERY 6 MONTHS FOR INSPECTION AND PREVENTIVE MAINTENANCE. ANNUAL FACTORY CALIBRATION CHECKS ARE STRONGLY RECOMMENDED TO VERIFY CONTINUED ACCURACY. THE DEVICE WAS SERVICED AND RETURNED TO THE CUSTOMER.
IT WAS REPORTED THAT THE ZIMMER AIR DERMATOME UNIT WAS CUTTING DIAMOND SHAPES, AS STATED BY THE SURGEON. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THE INCIDENT.