MDR report key: 2102743 · Received May 23, 2011

Report

Report Number
1720753-2011-07537
Event Type
Malfunction
Date Received
May 23, 2011
Date of Event
May 11, 2011
Report Date
May 23, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE PS1 POWER SUPPLY IN THE MAINFRAME WAS INCREASED FROM 4.8 VDC TO 5.25 VDC ON THE GENERATOR INTERFACE PRINTED CIRCUIT BOARD. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT DURING A PROCEDURE, THE SYSTEM DISPLAYED CONTROL PANEL AND COMMUNICATION ERROR MESSAGES. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900