FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2102722 · Received May 23, 2011

Report

Report Number
1720753-2011-07533
Event Type
Malfunction
Date Received
May 23, 2011
Date of Event
May 3, 2011
Report Date
May 23, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE GENERATOR INTERFACE PRINTED CIRCUIT BOARD, THE REAL TIME OPERATION SYSTEM AND THE GENERAL PURPOSE OPERATING SYSTEM WERE REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT INTERMITTENTLY THE SYSTEM WOULD NOT WORK WHEN IN CINE MODE. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1