FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2102722
·
Received May 23, 2011
Report
- Report Number
- 1720753-2011-07533
- Event Type
- Malfunction
- Date Received
- May 23, 2011
- Date of Event
- May 3, 2011
- Report Date
- May 23, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE GENERATOR INTERFACE PRINTED CIRCUIT BOARD, THE REAL TIME OPERATION SYSTEM AND THE GENERAL PURPOSE OPERATING SYSTEM WERE REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT INTERMITTENTLY THE SYSTEM WOULD NOT WORK WHEN IN CINE MODE. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |