FDA Adverse Event
Malfunction
Summary report: N
6800
MDR report key: 2102698
·
Received May 23, 2011
Report
- Report Number
- 1720753-2011-07515
- Event Type
- Malfunction
- Date Received
- May 23, 2011
- Date of Event
- May 11, 2011
- Report Date
- May 23, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MT, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REP ADJUSTED THE 5VDC VOLTAGE AT THE GENERATOR CABINET AND THE ELECTRONIC CABINET. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD LOCK UP DURING FLUOROSCOPIC X-RAY. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 6800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 6800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |