PARADIGM INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2011-81555
- Event Type
- Malfunction
- Date Received
- May 20, 2011
- Date of Event
- March 28, 2011
- Report Date
- May 17, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- LZG
- PMA / PMN Number
- K053177
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NOT APPLICABLE
Narratives
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
THE INSULIN PUMP FAILED DISPLACEMENT TEST. MOTOR ERROR ALARM DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE.
IT WAS REPORTED THAT A MOTOR ERROR ALARM OCCURRED DURING REWIND. THE PHYSICIAN STATED THAT HE COULD NOT OPERATE THE INSULIN PUMP. THE PHYSICIAN THEN STATED THAT HE RESTED THE INSULIN PUMP FOR FIVE MINUTES, CHANGED THE BATTERY, AND THAT HE STILL COULD NOT REWIND THE INSULIN PUMP. THE PHYSICIAN FURTHER STATED THAT THE MOTOR ERROR ALARM OCCURRED AGAIN AND REPEATED SEVERAL TIMES. TROUBLESHOOTING WAS PERFORMED AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-712EWS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |