INRATIO
Report
- Report Number
- 2027969-2011-01123
- Event Type
- Malfunction
- Date Received
- May 20, 2011
- Date of Event
- April 25, 2011
- Report Date
- May 20, 2011
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GJS
- PMA / PMN Number
- K021923
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST RESULT(S) WITH LAB RESULT(S) PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: (B)(6) 2011, INRATIO: 3.0, REFERENCE: 2.4, MEAN: 2.7, CONFIDENCE LIMIT: 1.7 - 3.8, RESULT: PASS. REPORTED RESULTS ARE WITHIN THE CONFIDENCE LIMITS FOR PASSING ACCURACY COMPARISON. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. IN-HOUSE THERAPEUTIC DONOR TESTING HAS BEEN PERFORMED ON REPORTED LOT 247451 ON (B)(4) 2011. RESULTS AS FOLLOWS: INRATIO: 1.6, INRATIO: 1.6, INRATIO: 1.6, REFERENCE: 1.64, BIAS THRESHOLD: 1.14 - 2.14; 2.7, 2.8, 2.7, 2.44, 1.44 - 3.44. THE MINIMUM TWO OUT OF THREE REPLICATES FOR EACH DONOR ARE WITHIN THE ACCEPTABLE BIAS FOR ACCURACY. PRODUCT PERFORMED AS EXPECTED. METER IS EXPECTED TO BE RETURNED AND WILL BE INVESTIGATED WHEN REC'D. (B)(4). ACTION THRESHOLD (B)(4) HAS REACHED. STRIP LOT IN COMPLAINT HAS STRIP CODE 62935 WHICH BRICK LOT NUMBER 237418 WAS ASSIGNED AND PACKAGED INTO TWO STRIP LOTS (247450 AND 247451). (B)(4). DUE TO THIS LOW OCCURRENCE RATE, BELOW THE TOTAL COMPLAINT ACTION THRESHOLD OF (B)(4). CORRECTIVE ACTION IS NOT REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.
"CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS:" DATE: (B)(6) 2011, INRATIO: 3.0, LAB: 2.4, TARGET RANGE IS 2.5-3.5.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC. | 100071 | 247451 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |