MULTITASK OPERATING SYSTEM FOR AUTOMIX COMPOINDER
Report
- Report Number
- 210256
- Event Type
- Death
- Date Received
- February 15, 1999
- Date of Event
- December 31, 1998
- Report Date
- February 11, 1999
- Manufacturer
- CLINTEC NUTRITION CO.
- Product Code
- LHI
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- KS, US
- Reporter Occupation
- RISK MANAGER
Narratives
AN ORDER READING 1/2 NORMAL SALINE WITH 1/2 U/CC HEPARIN AT 1CC/HOUR WAS REC'D BY PHARMACY. THE SOLUTION WAS COMPOUNDED USING THE MULTITASK OPERATING SYSTEM. SODIUM CHLORIDE 77MEQ WAS ENTERED INTO THE SYSTEM FOR COMPOUNDING A 24 ML BAG (WITH 100 ML OVERFILL). PROBLEMS IDENTIFIED INCLUDED LACK OF CLEARLY DISTINGUISHABLE SCREENS FOR VITAL DRUG ALERTS, POOR LABEL FORMAT AND CLARITY OF LABELS, COMPLEX ORDER ENTRY PROCESS, SCREENS DEPICTING SOLUTION AND ADDITIVE ASSIGNED STATIONS AND ID THAT APPEAR AT THE END OF ORDER ENTRY ALONG WITH WARNING SCREENS, CLINICALLY INSIGNIFICANT WARNINGS APPEAR AS FREQUENTLY AS VITAL WARNINGS, NO ACTION REQUIRED BY OPERATOR, SUCH AS PASSWORD OR OTHER RESPONSE TO ASSURE THAT VITAL WARNING IS ACKNOWLEDGED.
AN ORDER READING 1/2 NORMAL SALINE WITH 1/2 U/CC HEPARIN AT 1CC/HOUR WAS REC'D BY PHARMACY. THE SOLUTION WAS COMPOUNDED USING THE MULTITASK OPERATING SYSTEM. SODIUM CHLORIDE 77MEQ WAS ENTERED INTO THE SYSTEM FOR COMPOUNDING A 24 ML BAG (WITH 100 ML OVERFILL). PROBLEMS IDENTIFIED INCLUDED LACK OF CLEARLY DISTINGUISHABLE SCREENS FOR VITAL DRUG ALERTS, POOR LABEL FORMAT AND CLARITY OF LABELS, COMPLEX ORDER ENTRY PROCESS, SCREENS DEPICTING SOLUTION AND ADDITIVE ASSIGNED STATIONS AND ID THAT APPEAR AT THE END OF ORDER ENTRY ALONG WITH WARNING SCREENS, CLINICALLY INSIGNIFICANT WARNINGS APPEAR AS FREQUENTLY AS VITAL WARNINGS, NO ACTION REQUIRED BY OPERATOR, SUCH AS PASSWORD OR OTHER RESPONSE TO ASSURE THAT VITAL WARNING IS ACKNOWLEDGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MULTITASK OPERATING SYSTEM FOR AUTOMIX COMPOINDER | COMPUTER OPERATING SYSTEM (SOFTWARE) | LHI | CLINTEC NUTRITION CO. | VERSION 2.3 'E' | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 9 DAY | Death | ||
| 2 | 9 DAY | Death |