FDA Adverse Event Malfunction Summary report: N

WIRE COLLET

MDR report key: 2102413 · Received May 19, 2011

Report

Report Number
1811755-2011-01841
Event Type
Malfunction
Date Received
May 19, 2011
Date of Event
April 29, 2011
Report Date
April 29, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
GET
PMA / PMN Number
K943323
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THERE WAS DEBRIS BUILT UP INSIDE THE DEVICE. THIS CONDITION COULD CAUSE THE DEVICE TO NOT RETAIN THE WIRE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD NOT RETAIN A WIRE DURING ROUTINE MAINTENANCE. THERE WAS NO PATIENT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WIRE COLLET GET STRYKER INSTRUMENTS KALAMAZOO 08003

Patients

Seq Age Sex Outcome Treatment
1 UNK