FDA Adverse Event Malfunction Summary report: N

PULSE GEN MODEL 104

MDR report key: 2102394 · Received May 20, 2011

Report

Report Number
1644487-2011-01057
Event Type
Malfunction
Date Received
May 20, 2011
Date of Event
November 21, 2007
Report Date
April 21, 2011
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF PROGRAMMING HISTORY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PHYSICIAN FELT THAT THE PT'S VNS GENERATOR "INTENSIFIED F/U INDICATOR (IFI)" HAD TRIPPED EARLIER THAN EXPECTED. BASED ON THE PT'S SETTINGS AND THE BATTERY LIFE ESTIMATION TABLES IN THE PHYSICIAN'S MANUAL, THE VNS GENERATOR IFI INDICATOR LIKELY SHOULD NOT HAVE TRIPPED. THE ALERT IS MEANT TO SERVE AS A NOTIFICATION TO THE TREATING PHYSICIAN THAT MORE FREQUENT CLINICAL MONITORING IS RECOMMENDED AS THE GENERATOR BATTERY WAS MEASURED TO BE AT A LEVEL WHERE MORE FREQUENT CLINICAL MONITORING IS RECOMMENDED. A REVIEW OF THE PREVIOUS PROGRAMMING HISTORY IDENTIFIED THAT THE VOLTAGE MEASURED BY THE GENERATOR LESS THAN ONE YEAR AFTER IMPLANT (2.756V) WAS UNCHARACTERISTICALLY LOW BASED ON PT THERAPY SETTINGS, SYSTEM IMPEDANCE AND IMPLANT DURATION. ADDITIONALLY, BASED ON THE TIMING OF THE INITIAL REPORT, IT APPEARS THAT THE MEASURED GENERATOR VOLTAGE HAS BEEN WITHIN THE IFI LIMITS (WHICH CORRESPONDS TO THE LAST 25% OF REMAINING BATTERY CAPACITY) FOR APPROX 2 YEARS, WHICH IS UNEXPECTED BEHAVIOR AT THESE VOLTAGES. A REVIEW OF THE GENERATOR'S DEVICE HISTORY RECORD (DHR) REVEALED THAT ALL ELECTRICAL TESTS/VISUAL INSPECTIONS WERE PASSED PRIOR TO DEVICE SHIPMENT AND THAT THE VOLTAGE MEASUREMENT ACCURACY OF THE GENERATOR WAS NEAR THE LOWER LIMIT OF THE SPEC, WHICH APPEARS TO EXPLAIN THE LOWER MEASURED VOLTAGE VALUE. ATTEMPTS FOR ADDITIONAL INFO ARE IN PROGRESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PULSE GEN MODEL 104 LYJ CYBERONICS, INC. 104 1952

Patients

Seq Age Sex Outcome Treatment
1 56 YR Female