SYRINGE 0.5ML 31G TW 6MM BLS 400 SG
Report
- Report Number
- 3024508819-2024-00512
- Event Type
- Malfunction
- Date Received
- December 26, 2024
- Report Date
- February 13, 2025
- Manufacturer
- EMBECTA MEDICAL I LLC - HOLDREGE, NE
- Product Code
- FMF
- PMA / PMN Number
- K992734
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ADDITIONAL INFORMATION WAS ADDED TO: B4, G6, H2, H3, H11. CORRECTION TO: H6 (COMPONENT CODE, TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS). INVESTIGATION SUMMARY: NO SAMPLES WERE RETURNED THEREFORE THE INVESTIGATION WAS PERFORMED BASED ON THE PHOTO(S) PROVIDED. EMBECTA WAS ABLE TO CONFIRM THE CUSTOMER-INDICATED ISSUE. THIS IS THE 1ST COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND NO CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED AT THIS TIME.
MATERIAL # 328447. MATERIAL DESCRIPTION - SYRINGE 0.5ML 31G TW 6MM BLS 400 SG. MATERIAL LOT# - 4103061. COMPLAINT DESCRIPTION ¿ OUR NURSES AT (B)(6) HOSPITAL IN (B)(6), NY ARE SEEING A CONCERN WITH OUR CURRENT INSULIN SYRINGE, REF# (B)(4), LOT# 4103061P1. PLEASE SEE THE ATTACHED PHOTOS OF WHAT APPEARS TO BE DOUBLE INK MARKS FOR UNIT MARKINGS. THIS PRESENTS A SAFETY ISSUE FOR DOSING. WANTED TO MAKE YOU AWARE THAT ONE OF OUR NURSES NOTICED SEVERAL OF OUR INSULIN NEEDLES HAVE A FAINT DOUBLE PRINT ON THEM, WHICH COULD BE DANGEROUS FOR INSULIN ADMINISTRATION. I HAVE A FEW OF THEM IN MY OFFICE WITH THE PACKAGING IF YOU NEED THEM. NOTES (IF ANY) ¿ ATTACHING THE PHOTOS THAT CUSTOMER HAS PROVIDED FOR FURTHER REFERENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2510599 | SYRINGE 0.5ML 31G TW 6MM BLS 400 SG | SYRINGE, PISTON | FMF | EMBECTA MEDICAL I LLC - HOLDREGE, NE | 328447 | 4103061 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Prefer Not To Disclose |