Narratives

Description of Event or Problem · 1

DURING MFR REVIEW OF A VNS PT'S PROGRAMMING HISTORY, IT WAS IDENTIFIED THAT ON (B)(6) 2008 THAT THE PARAMETERS WERE DEFAULT SYSTEMS DIAGNOSTICS SETTING (0MA/20HZ/500PULSEWIDTH/30 SEC ON/60 MIN OFF). THE SETTINGS WERE CORRECTED TO 2MA/20HZ/250PULSEWIDTH/30 SEC ON/1.1 MIN OFF. IT IS SUSPECTED AN INTERRUPTED SYSTEMS DIAGNOSTICS TEST OCCURRED WITH A DIFFERENT VNS PROGRAMMING SYSTEM AT THE PREVIOUS OFFICE VISIT ON (B)(6) 2008.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. 250

Patients

Seq Age Sex Outcome Treatment
1 18 YR