THYROTROPIN
Report
- Report Number
- 1823260-2011-02798
- Event Type
- Malfunction
- Date Received
- May 25, 2011
- Date of Event
- May 3, 2011
- Report Date
- May 25, 2011
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JLW
- PMA / PMN Number
- K961491
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
THE CUSTOMER STATED THEY RECEIVED A QUESTIONABLY LOW THYROTROPIN (TSH) RESULT ON THEIR COBAS E411 RACK SYSTEM (SERIAL NUMBER (B)(4)). THE TECHNOLOGIST QUESTIONED THE LOW RESULT AND DECIDED TO REPEAT THE TEST. THE REPEAT TESTING WAS PERFORMED ON THE SAME E411 ANALYZER. THE PATIENT'S INITIAL TSH RESULT WAS 0.034 UIU/ML ACCOMPANIED BY A DATA FLAG. THE FIRST REPEAT RESULT WAS 4.29 UIU/ML. THE SECOND REPEAT RESULT WAS 4.32 UIU/ML. THE ORIGINAL ERRONEOUS RESULT WAS NEVER REPORTED OUTSIDE THE LABORATORY SO THE PATIENT WAS NOT ADVERSELY AFFECTED AND THERE WAS NO RESULTING TREATMENT. THE CUSTOMER REFUSED A FIELD SERVICE VISIT. UPON FURTHER INVESTIGATION, NO ROOT CAUSE COULD BE DETERMINED. CALIBRATION AND QUALITY CONTROL BOTH PASSED PRIOR TO THE ANALYSIS INDICATING THE CAUSE WAS LIKELY NOT A REAGENT MALFUNCTION. THE CUSTOMER REPORTED SEEING A DROP OF LIQUID ON THE INNER WALL OF THE SAMPLE TUBE. THE DROP OF LIQUID COULD BE A POSSIBLE REASON FOR PREMATURE LIQUID LEVEL DETECTION LEADING TO AN INITIAL FALSE LOW RESULT. THERE WERE NO DETAILED PATIENT DATA PROVIDED IN THIS CASE. IT WAS UNKNOWN WHY DETERMINATION OF TSH WAS PERFORMED. THE LOW RESULT WAS NOT REPORTED OUTSIDE OF THE LAB. THE CUSTOMER CONFIRMED THAT THE PATIENT WAS NOT ADVERSELY AFFECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THYROTROPIN | RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE | JLW | ROCHE DIAGNOSTICS | NA | 16107903 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 050 YR |