FDA Adverse Event
Malfunction
Summary report: N
INFINITI VISION SYSTEM OZIL
MDR report key: 2102242
·
Received May 19, 2011
Report
- Report Number
- 2028159-2011-00568
- Event Type
- Malfunction
- Date Received
- May 19, 2011
- Date of Event
- April 19, 2011
- Report Date
- April 19, 2011
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CENTER
- Product Code
- HQC
- PMA / PMN Number
- K082845
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY REP EXAMINED THE SYSTEM AND REPLACED THE FOOTSWITCH CABLE. PREVENTATIVE MAINTENANCE WAS PERFORMED AND THE SYSTEM WAS THEN TESTED MET PRODUCT SPECS. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY WHEN ADD'L REPORTABLE INFO BECOMES AVAILABLE. (B)(4).
Description of Event or Problem · 1
A NURSE REPORTED THE SYSTEM WOULD NOT RECOGNIZE THE FOOTSWITCH DURING SURGERY. THE FOOTSWITCH WAS SWITCHED OUT AND THE SURGERY WAS COMPLETED WITH NO HARM OR INJURY TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFINITI VISION SYSTEM OZIL | PHACOFRAGMENTATION SYSTEM | HQC | ALCON - IRVINE TECHNOLOGY CENTER | INFINITI OZIL | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |