FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2102169 · Received May 19, 2011

Report

Report Number
1720753-2011-07481
Event Type
Malfunction
Date Received
May 19, 2011
Date of Event
May 5, 2011
Report Date
May 19, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE LEFT MONITOR AND THE COOLING FANS WERE REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM'S LEFT MONITOR WOULD NOT START UP AND THE COOLING FANS WERE FAULTY. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1