FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2102168 · Received May 19, 2011

Report

Report Number
1720753-2011-07467
Event Type
Malfunction
Date Received
May 19, 2011
Date of Event
May 9, 2011
Report Date
May 19, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE FLUORO FUNCTIONS PRINTED CIRCUIT BOARD AND POWER SIGNAL INTERFACE PRINTED CIRCUIT BOARD WERE REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE MAINFRAME SYSTEM SHUT DOWN OUTSIDE OF A CASE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1