FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2102161
·
Received May 19, 2011
Report
- Report Number
- 1720753-2011-07471
- Event Type
- Malfunction
- Date Received
- May 19, 2011
- Date of Event
- May 9, 2011
- Report Date
- May 19, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE HARD DRIVE WAS REPLACED AND THE SOFTWARE WAS RELOADED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SVC.
Description of Event or Problem · 1
THE CUSTOMER REPORTED DURING A CASE THE SYSTEM LOCKED UP THEN WOULD NOT BOOT UP. THE CASE WAS COMPLETED WITH ANOTHER C-ARM SYSTEM. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |