FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2102156 · Received May 19, 2011

Report

Report Number
1720753-2011-07460
Event Type
Malfunction
Date Received
May 19, 2011
Date of Event
May 9, 2011
Report Date
May 19, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE UPPER STEERING UNIVERSAL JOINT WAS CLEANED AND LUBRICATED AND THE STEERING ASSEMBLY WAS REASSEMBLED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SVC.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM HAD A PROBLEM WITH THE STEERING ON THE C-ARM OUTSIDE OF A CASE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1