FDA Adverse Event Malfunction Summary report: N

6600

MDR report key: 2102130 · Received April 20, 2011

Report

Report Number
1720753-2011-06371
Event Type
Malfunction
Date Received
April 20, 2011
Date of Event
April 6, 2011
Report Date
April 20, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE CONNECTORS WERE LOOSE THEREFORE THE VIDEO CABLE WAS REPLACED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE 6600 SYSTEM WAS NOT CAPTURING IMAGES CORRECTLY. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6600

Patients

Seq Age Sex Outcome Treatment
1