FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2102122 · Received April 20, 2011

Report

Report Number
1720753-2011-06377
Event Type
Malfunction
Date Received
April 20, 2011
Date of Event
April 5, 2011
Report Date
April 20, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MFR'S SERVICE REP PERFORMED AN ON-SITE INVESTIGATION. THE SERVICE REP REPLACED THE BRAKE PAD, THE X-RAY TUBE COVER, AND RESEATED THE GENERATOR INTERFACE BOARD. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9800 SYSTEM WOULD NOT SAVE IMAGES AND THE SCREEN WENT BLACK DURING FLUOROSCOPY. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1